The CS451 is a pressure transducer for water-level measurements in canals, wells, ponds, harbors, lakes, streams and tanks. It has a stainless-steel case that can be submerged in most canals, wells, ponds, lakes, and streams. The CS451 outputs either a digital SDI-12 or RS-232 signal to indicate observed pressure and temperature. This output can be read by many of our dataloggers.
The CS451 replaces the CS450 transducer. The new transducers have a smaller gap between the water ports and the diaphragm so that less air is trapped that the user must remove during deployment. Trapped air causes the transducer's readings to drift as the air slowly dissolves into the water.
Read MoreThe CS451 consists of a piezoresistive sensor and a temperature sensor housed in a 316L stainless-steel case. It has a rugged Hytrel cable that remains flexible, even under harsh environmental conditions. The cable incorporates a vent tube to compensate for atmospheric pressure fluctuations. The vent tube terminates inside a desiccant tube, which prevents water vapor from entering the inner cavity of the transducer. The sensor ships with a desiccant tube that can be replaced in the field.
The CS451 has several pressure range options and two accuracy options (see Ordering Info). The standard accuracy option provides ±0.1% FS TEB over the 0° to 60°C temperature range. The high accuracy option provides ±0.05% FS TEB over the 0° to 60°C temperature range and includes a calibration certificate. TEB is the combined errors due to nonlinearity, hysteresis, non-repeatability, and thermal effects over the compensated temperature range, per ISA S51.1. Please note that the high accuracy option is not available for the 0 to 2.9 psig range option.
Campbell Scientific offers the A150 Desiccated Case that allows the CS451 to be connected to a prewired enclosure (see Ordering Information).
Measurement Time | < 1.5 s |
Output Options | SDI-12 (version 1.3) 1200 bps; RS-232 9600 bps |
Water-Level Resolution | 0.0035% FS |
Worst-Case Temperature Resolution | 0.006°C |
Dry Storage Temperature Range |
-40° to +100°C WARNING: Sensor could be damaged if encased in frozen ice. |
Operating Temperature Range |
0° to 60°C WARNING: Sensor could be damaged if encased in frozen ice. |
Temperature Accuracy | ±0.2°C |
Overpressure | 2 x pressure range |
Power Requirements | 6 to 18 Vdc |
Cable Type | Hytrel Jacket, five conductor, 26 AWG |
NPT Fitting | 1/4-in. NPS |
Top Cone Material | Delrin |
Body Material | 316L stainless steel |
Element Material | 316L stainless steel |
Distance |
|
Ingress Protection | Exceeds IP68 |
Diameter | 21.34 mm (0.84 in.) |
Cable Outer Diameter |
|
Length | 213.36 mm (6.875 in.) |
Cable Weight | 0.0421 kg/m (0.0283 lb/ft) |
Weight | 0.17 kg (0.37 lb) without cable |
Air Gap |
|
Standard & Weighted Nose Cone | 0.653 cm (0.257 in.) |
NPT Fitting | 2.72 cm (1.07 in.) |
Power Consumption |
|
Quiescent | < 50 µA |
Measurement/Communication | 8 mA (1 s measurement) |
Maximum | 40 mA |
Measurement Ranges at Fresh Water Depths |
|
0 to 2.0 m (6.7 ft) |
|
0 to 5.1 m (16.7 ft) |
|
0 to 10.2 m (33.4 ft) |
|
0 to 20.4 m (67 ft) |
|
0 to 50.9 m (167 ft) |
|
0 to 102 m (334.5 ft) |
|
Accuracy |
|
Standard Accuracy Option |
±0.1% full-scale-range TEB Total Error Band (TEB) includes the combined errors due to nonlinearity, hysteresis, nonrepeatability, and thermal effects over the compensated temperature range, per ISA S51.1. |
High Accuracy Option |
±0.05% full-scale-range TEB The high accuracy (±0.05% full-scale range) option is not available for the 0 to 2.9 psig range option. Total Error Band (TEB) includes the combined errors due to nonlinearity, hysteresis, nonrepeatability, and thermal effects over the compensated temperature range, per ISA S51.1. |
Maximum Cable Length |
|
SDI-12 |
|
RS-232 | 60 m (200 ft) |
Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.
Includes the current firmware for the CS451 and CS456.
Number of FAQs related to CS451: 6
Expand AllCollapse All
The CS45x-series pressure transducers are suitable for monitoring lake levels. However, be sure to submerge the pressure transducer below the frost line. If ice forms against the diaphragm, it will deform the diaphragm and cause permanent damage.
No. Because of the SDI-12 protocol, the CS45x-series pressure transducers can only make measurements every 1.5 seconds.
Yes, but a data logger is required. The CS45x-series pressure transducers are not autonomous. They require a data logger to initiate and retrieve a measurement. The data logger can initiate an event-based measurement.
In contrast, the CRS45x-series water-level recording sensors support event-based recording using built-in data-collection features.
No. A data logger must be used with the CS45x-series pressure transducers.
No. The A200 is only used to configure the CS451 and CS456 pressure transducers.